Servicing
Optional Solenoid Valves Removal
7-52
42iQLS Instruction Manual
Thermo Scientific
3.
Using a #1 Phillips drive, loosen two #4-40 pan head screws.
4.
Slide valve out.
Equipment Damage
Do not disconnect the plumbing from the valve end.
Disconnect from the attaching end only. This will prevent damaging and
leaks from the valve end.
▲
Figure 7–55.
Mode-Lag-Perm Hardware, Plumbing