Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-30
42iQLS Instruction Manual
Thermo Scientific
Figure 7–32.
Remove DMC PMT Cooler
Use the following procedure to replace the PMT cooler board.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove DMC Converter as described in “
PMT Cooler Shroud
Removal
” in this chapter.
3.
Unplug all electrical connections from the board.
4.
Using a #2 Phillips drive, unfasten six #6-32 pan head screws.
5.
Pull board outwards.
PMT Cooler Board
Replacement
Summary of Contents for 42iQLS