Servicing
Power Supply Replacement
Thermo Scientific
42iQLS Instruction Manual
7-23
Use the following procedure to replace the power supply.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug all electrical shown J9, J10, J24, J25, J26, and ground.
3.
Unfasten captive hardware.
4.
Slide power supply left clearing three case floor plate keyways.
Figure 7–25.
Removing Power Supply
5.
Pull power supply up.
6.
Replace power supply and reassemble in reverse order.
Power Supply
Replacement