Operation
Settings
Thermo Scientific
42iQLS Instruction Manual
3-73
The Preventive Maintenance screen reminds the user when certain
instrument components need to be serviced/replaced. When the “Months
Left” has decreased to 2, the row is highlighted yellow. If the “Months
Left” is 1 or less, the row is highlighted red and the maintenance icon
(gears) will appear in the status bar located at the bottom of the screen.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Maintenance>Preventive
Maintenance
The Preventive Maintenance screen contains the following information:
●
Across:
●
Component:
Device to be routinely serviced or replaced.
●
Interval in Months:
Expected period of time before a component
needs to be checked and/or serviced.
●
Months Left:
Count down timer in months. Remaining time since
the beginning of the service interval. When the value is 1 or less,
the row will be highlighted and it is suggested that the component
should be checked and/or serviced.
●
Reset:
Once the component is serviced/replaced, the user presses the
Reset button and the “Months Left” value resets to the “Service
Interval in Months” value.
●
Alert:
Allows the user to opt out of receiving preventive
maintenance alerts. Displays Enabled/Disabled for each
component.
Preventive Maintenance
Summary of Contents for 42iQLS
Page 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Page 2: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Page 192: ......
Page 296: ......
Page 304: ......
Page 316: ......
Page 317: ......