Servicing
Optional DMC Oxygen Sensor
7-60
42iQLS Instruction Manual
Thermo Scientific
Use the following to replace the oxygen sensor.
Equipment required:
Hex drive, 3mm
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-53
.
2.
Unplug oxygen sensor cable.
Figure 7–64.
Replacing the Oxygen Sensor
3.
Disconnect tubing.
4.
Using a 3mm hex drive, unfasten three m4 socket cap head screws.
5.
Slide out oxygen sensor.
6.
Replace oxygen sensor and reassemble in reverse order.
Oxygen Sensor
Replacement