Servicing
DMC Ozonator
Thermo Scientific
42iQLS Instruction Manual
7-47
Figure 7–49.
Flow Switch Remove
Use the following procedure to replace the ammonia scrubber.
Equipment Required:
Phillips drive, #2
1.
Remove DMC Ozonator as described in “
DMC Ozonator
” in this
chapter.
2.
Disconnect plumbing 2X.
3.
Pull Ammonia scrubber off clips 2X.
Ammonia Scrubber
Replacement