Operation
Settings
3-52
42iQLS Instruction Manual
Thermo Scientific
●
Chamber Temp Sensor Open:
Displays OK/Fail for chamber
temperature sensor open.
●
Cooler Temperature:
Displays OK/Fail for cooler temperature. User
can adjust low and high alarm limits.
●
Cooler Temp Sensor Short:
Displays OK/Fail for cooler temperature
sensor short.
●
Cooler Temp Sensor Open:
Displays OK/Fail for cooler temperature
sensor open.
●
Cooler Current:
Displays the cooler current reading.
●
Cooler Current too Low:
Displays OK/Fail for cooler current too
low.
●
Cooler Current too High:
Displays OK/Fail for cooler current too
high.
●
Cooler Voltage:
Displays the current cooler voltage reading.
●
Cooler Voltage too Low:
Displays OK/Fail for cooler voltage too low.
●
Cooler Voltage too High:
Displays OK/Fail for cooler voltage too
high.
●
PMT Voltage:
Displays the current PMT voltage reading.
●
PMT Voltage too Low:
Displays OK/Fail for PMT voltage too low.
●
PMT Voltage too High:
Displays OK/Fail for PMT voltage too
high.
●
Frequency:
Displays the current frequency reading.
●
Frequency too Low:
Displays OK/Fail for frequency too low.
●
Frequency too High:
Displays OK/Fail for frequency too high.
●
Low Gain:
Displays the current low gain reading.
●
Gain:
Displays the current gain reading.
●
Board Communication:
Displays OK/Fail for board communication
status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
3.3 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
5 V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
Summary of Contents for 42iQLS
Page 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Page 2: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Page 192: ......
Page 296: ......
Page 304: ......
Page 316: ......
Page 317: ......