Operation
Settings
3-64
42iQLS Instruction Manual
Thermo Scientific
The Valve and Pump Resets screen allows the user to reset a valve or pump
after a failure due to excessive amperage.
Note
This symbol denotes that the device needs to be reset.
▲
Note
Resetting one valve will reset all valves.
▲
Home Screen>Settings>Health Check>Status and Alarms>Valve and
Pump Resets
The Valve and Pump Resets screen contains the following information:
●
Sample Valve Reset:
Resets sample valve.
●
Zero Valve Reset:
Resets zero valve.
●
Span Valve Reset:
Resets span valve.
●
NO/NO
x
Valve Reset:
Resets NO/NO
x
valve.
●
Pump Reset:
Resets pump.
Valve and Pump Resets