Servicing
Optional DMC Oxygen Sensor
Thermo Scientific
42iQLS Instruction Manual
7-61
Use the following to install the oxygen sensor DMC.
Equipment required:
Phillips drive, #2
1.
Slide oxygen sensor assembly downwards.
Note
Make sure oxygen sensor plate goes over the partition panel keyway
stud.
▲
Figure 7–65.
Installing the Oxygen Sensor DMC
2.
Slide oxygen sensor assembly right.
Note
Make sure oxygen sensor plate fits under the stud.
▲
Note
Make sure oxygen sensor plate slides into side pocket.
▲
3.
Using a #2 Phillips drive, fasten captive hardware.
Oxygen Sensor DMC
Install