Servicing
Optional DMC Oxygen Sensor
7-56
42iQLS Instruction Manual
Thermo Scientific
Use the following procedure to replace the oxygen sensor board.
Equipment required:
Torque screw driver, T15 or slot screw driver, 3/16
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-53
.
2.
Unplug J1.
3.
Using a T15 torque drive, unfasten two captive hardware.
Figure 7–60.
Replacing the Oxygen Sensor DMC Board
4.
Slide board upwards.
Note
Clear keyway.
▲
Oxygen Sensor
Board Replacement