Operation
Calibration
Thermo Scientific
42iQLS Instruction Manual
3-33
The Point 1 screen allows the user to view and set the selected calibration
point. The point 2 screen functions the same way. Therefore, the following
example of point 1 screen applies to point 2 as well.
Home Screen>Calibration>Advanced Calibration>2-Point O
2
Calibration>Continue to 2-Point O
2
Calibration>Point 1
The Point 1 screen contains the following information:
●
Edit O
2
Span Concentration:
User enters O
2
span concentration.
●
Current O
2
Concentration:
Read only. After selecting the value on the
Edit O
2
Concentration button and after pressing the Calibrate O
2
button, the Point 1 Concentration value is updated.
●
Calibrate O
2
:
Calibrates point 1 coefficient.
Point 1
2