Servicing
Step POL Board Replacement
7-26
42iQLS Instruction Manual
Thermo Scientific
Figure 7–28.
Clear Partition Keyway Step POL Board
6.
Slide step board 1 upwards clearing the partition panel keyway.
7.
If replacing step board 1, make sure switch 1 and 2 are pointed away
from ON (
Figure 7–29
). If replacing optional step board 2, make sure
switch 1 is pointed towards ON and switch 2 is pointed away from ON
(
Figure 7–29
).
8.
Replace Step POL board and reassemble in reverse order.