Operation
Settings
Thermo Scientific
42iQLS Instruction Manual
3-97
The Compensation screen provides compensation for any changes to the
instrument’s output signal due to internal instrument temperature,
pressure, and oxygen variations (if enabled).
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
●
Temp Compensation:
Toggles temperature compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on
the analyzer's subsystems and output have been empirically determined.
This empirical data is used to compensate for any changes in
temperature.
●
Pressure Compensation:
Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to bench pressure variations. The effects
of bench pressure changes on the analyzer's subsystems and output have
been empirically determined. This empirical data is used to compensate
for any change in chamber pressure.
●
O
2
Correction:
Allows for the correction of the NO, NO
2
, and NO
x
concentrations based on O
2
readings.
●
Standard Temperature for Gas Units:
Allows the user to enter a value
between 0 and 30
°
C to correct the
g/m
3
, mg/m
3
, and g/m
3
concentration calculations.
Compensation
Summary of Contents for 42iQLS
Page 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Page 2: ......
Page 18: ......
Page 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Page 192: ......
Page 296: ......
Page 304: ......
Page 316: ......
Page 317: ......