Operation
Settings
Thermo Scientific
42iQLS Instruction Manual
3-65
The Analog I/O screen provides status and alarms related to the analog
input/output module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and
Alarms>More>Analog I/O
The Analog I/O screen contains the following information:
●
Across:
●
Analog IO:
This column lists items associated with the analog I/O.
●
Value:
Displays the current value for each item.
●
Low Alarm:
Displays low alarm status for each item.
●
High Alarm:
Displays high alarm status for each item.
●
Units:
Displays units for each item.
●
Down:
●
Voltage Output Channel 1–6:
Displays real-time voltage output for
each channel.
●
Current Output Channel 1–6:
Displays real-time current output for
each channel.
●
Chip Temperatures:
Displays OK/Fail for chip temperatures.
●
Chip 1–3 Communication:
Displays OK/Fail for each chip
communication.
●
Test Mode:
Displays test mode on or off.
Analog I/O