Operating Inform ation-2445 Service
ACTimiATjOli A
■
■
IP
T
A
A
C
A
IT
H O I IV J IV I/-A I i o i v i H U J U O I I V I C I V I
1.
Preset instrument controls and obtain a baseline
trace as described in "In itia l Setup." Set 20 MHz BW
L IM IT On (in)
2.
Connect a 10X probe to the CH 1 OR X input
connector and connect the probe tip to the CALIBRATOR
output.
3.
Adjust the Channel 1 POSITION control to center
the display on the screen.
4. Set A and B SEC/DIV controls at 1 jus.
n no ro tin n
\A/i + hin nominal limitc Tho Ho halanno
attained by the Auto DC Balance adjustment remains
valid as long as the instrument is operating w ithin 5°C o f
the ambient temperature at which the adjustment was
performed provided a 20-minute warm-up period is allowed
before performing the adjustment. To initiate the adjust
ment, set both the Channel 1 and Channel 2 input coupling
switches to AC. Then simultaneously push up on both
switches. An alternate method o f entering the auto
adjustment mode is possible w ith only one of the input
coupling switches set to AC. Press up and hold the input
coupling switch that is not set to AC, then press up the
other input coupling switch. With either method, the
instrument w ill enter an auto-adjustment mode fo r about
ten seconds. When the Auto DC Balance adjustment cycle
is complete, the instrument w ill return to the normal
operating mode.
5.
Slowly adjust the FOCUS control to its optimum
setting (best-defined display).
NOTE
I f the AST/G adjustment is correctly set already,
a ll portions o f the trace w ill come into sharpest focus
at the same position o f the FOCUS control.
6.
If focusing is not uniform over the entire graticule
area, use a small-bladed screwdriver to adjust the ASTIG
control (see Figure 2-2).
NOTE
I f a circu it defect prevents accurate dc balance, the
routine halts and LIM IT is displayed. Push the Trigger
COUPLING switch up to continue balancing the
remainder o f the circuitry.
I f power to the instrument is interrupted before the
balancing cycle is completed, an error w ill be detected
by the next power-on self test. Press A /B TRIG to
e xit the diagnostic m onitor and restart the A uto
DC Balance adjustment to allow the cycle to be
completed.
7.
Since the ASTIG and FOCUS adjustments interact,
repeat steps 5 and 6 until the best-defined display over the
entire graticule area is obtained.
NOTE
Once it is set, the AST/G adjustment should be
correct fo r any display. However, it may be necessary
to reset the FOCUS control slightly when the
IN TEN SITY control setting is changed.
AUTO DC BALANCE ADJUSTMENT
The 2445 can automatically perform a dc-balance
adjustment o f Channel 1 and Channel 2. This adjustment
assures that the trace shifts associated w ith attenuator
stepping, changing the variable volts per division setting,
and switching Channel 2 between noninverted and Inverted
PROBE LOW-FREQUENCY
COMPENSATION
Misadjustment of probe compensation is one of the
sources o f measurement error. The attenuator probes
are equipped w ith compensation adjustments. To ensure
optimum measurement accuracy, always check probe
compensation before making measurements. Probe low-
frequency compensation is accomplished as follows:
1.
Preset instrument controls and obtain a baseline trace
as described in "In itia l Setup." Set 20 MHz BW LIM IT
On (in).
2.
Connect the two 10X probes (supplied w ith the
instrument) to the CH 1 OR X and the CH 2 input con
nectors. Observe that the CHANNEL 1 SCALE FACTOR
on the readout display changes from 10 mV to 100 mV
when the 10X p robe is attached.
2-18
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...