Maintenance— 2445 Service
Table 6-10
Pots and Switches Column
and Row Code Definitions
Row
Column
Definition
Row
Column
Definition
Code
Code
Code
Code
0
0
Trig COUPLING Down
5
0
B SEC/DIV LSB
0
1
Trig COUPLING Up
5
1
B SEC/DIV Bit 2
0
2
Unused
5
2
B SEC/DIV Bit 3
0
3
CH 1 Coupling Down
5
3
B SEC/DIV Bit 4
0
4
CH 1 Coupling Up
5
4
B SEC/DIV MSB
1
0
CH 4 VOLTS/DIV
6
0
CH 1 VERT MODE
1
1
CH 3 VOLTS/DIV
6
1
CH 2 VERT MODE
1
2
Unused
6
2
ADD VERT MODE
1
3
CH 2 Coupling Down
6
3
CH 3 VERT MODE
1
4
CH 2 Coupling Up
6
4
CH 4 VERT MODE
2
0
CH 1 VOLTS/DIV LSB
7
0
Unused
2
1
CH 1 VOLTS/DIV Bit 2
7
1
B ENDS A
2
2
CH 1 VOLTS/DIV Bit 3
7
2
Unused
2
3
CH 1 VOLTS DIV MSB
7
3
CHOP/ALT
2
4
CH 2 INVERT
7
4
BW LIMIT
3
0
CH 2 VOLTS/DIV LSB
8
0
X I0 MAG
3
1
CH 2 VOLTS/DIV Bit 2
8
1
TRACKING/INDEP
3
2
CH 2 VOLTS/DIV Bit 3
8
2
A t
3
3
CH 2 VOLTS/DIV MSB
8
3
AV
3
4
Horiz Display Select
8
4
Trig SLOPE
4
0
A SEC/DIV LSB
9
0
Trig SOURCE Down
4
1
A SEC/DIV Bit 2
9
1
Trig SOURCE Up
4
2
A SEC/DIV Bit 3
9
2
Trig MODE Down
4
3
A SEC/DIV Bit 4
9
3
Trig MODE Up
4
4
A SEC/DIV MSB
9
4
A/B TRIG Select
EAROM EXAMINE (Exerciser 02).
This routine allows the
operator to examine the contents of any or all EAROM loca
tions. The EAROM has 100 (decimal) locations (63 hexadec
imal). Addresses above 63 (hex) are not defined. When
entered, the Exerciser displays the contents of EAROM lo
cation 00 (hex) on the top line of the crt display. Calibration
constants reside between addresses 01 (hex) and 4C (hex)
and each should have odd parity as explained below. The
remaining locations may be of either parity. The readout dis
play line has the following format:
AA DDDDP
The format is defined as follows:
“AA”
is the eight-bit address in hexadecimal notation.
“DDDD”
is the 14-bit word stored at that location (13 bits
of data and one parity bit).
“P”
is a parity indicator for the data word: X indicates
even parity; blank is odd parity.
Pushing the MODE switch up or down will increment or
decrement the EAROM address by 16 (10 hex) respectively.
Similarly, pushing the SOURCE switch up or down will in
crement or decrement the address by 1 respectively.
CYCLE ERROR CLEAR (Exerciser 03).
This routine pro
vides a way for the operator to clear the cycle-failure data
written to the EAROM when a CYCLE mode failure occurs.
Until the data is cleared, each time the instrument is
powered up, the Diagnostic Monitor is entered and a diago
nal line is display across the crt.
CYCLE mode, when entered by removing the CAL/NO
CAL jumper (P501), causes the instrument to continuously
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...