Performance Check— 2445 Service
c.
Set the generator output level for an 8-division display
of the reference signal on CH 1.
d.
Adjust either the CH 1 VAR control or the CH 2 VAR
control for a minimum ADD display amplitude while leaving
the other control in the calibrated detent (whichever pro
vides the best CMRR).
e. Set the generator frequency to 50 MHz.
f. Set the SEC/DIV switch to 20 ns.
g.
CHECK—ADD display amplitude is 0.4 division or
less (discount trace width).
h.
Set ADD and INVERT Off (buttons out) and rotate the
CH 1 and CH 2 VAR controls CW to their calibrated detent
positions.
i. Disconnect the test setup.
7. Check Channel Isolation.
a. Set:
VERTICAL MODE
CH 1, CH 2,
CH 3, and CH 4
CHOP/ALT
Input Coupling
CH 1 and CH 2
VOLTS/DIV
CH 1 and CH 2
CH 3 and CH 4
TRIGGER
SOURCE
A and B SEC/DIV
b.
Connect the Leveled Sine-Wave Generator to the CH
1 OR X input connector via a 50 Q BNC cable.
c.
Set the generator frequency to 100 MHz and adjust
the output level for an 8-division display.
d. CHECK—Amplitude of each trace other than CH 1 is
0.08 division or less (discount trace width).
e.
Move the signal to the CH 2 input connector and
change the TRIGGER SOURCE switch to CH 2.
f. CHECK— Amplitude of each trace other than CH 2 is
0.08 division or less (discount trace width).
g. Add a 50
Q
BNC termination to the BNC cable and
move the signal to CH 3.
h. Set the TRIGGER SOURCE switch to CH 3 and ad
just the generator output for a signal display amplitude of 8
divisions.
i. CHECK— Amplitude of each trace other than CH 3 is
0.16 division or less (discount trace width).
j.
Move the signal to CH 4 input connector and set
TRIGGER SOURCE to CH 4.
k. CHECK— Amplitude of each trace other than CH 4 is
0.16 division or less (discount trace width).
I.
Connect the generator to the CH 1 OR X input
connector.
m. Set the TRIGGER SOURCE switch to CH 1.
n.
Set the generator output frequency to 150 MHz and
the output level for an 8-division display.
o. CHECK— Amplitude of each trace other than CH 1 is
0.16 division or less (discount trace width).
p. Move the signal to the CH 2 input connector and set
the TRIGGER SOURCE switch to CH 2.
q. fcHECK—Amplitude of each trace other than CH 2 is
0.16 division or less (discount trace width).
r. Disconnect the test setup.
On (buttons in)
ALT (button out)
50 0 DC
0.1 V
0.1 V (buttons out)
CH 1
20 ns (knobs locked)
4-11
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...