Adjustment Procedure— 2445 Service
q. CHECK— Readout
indicates
CH
4,
2V,
and
(step) 221.
r. Change the generator output level to 2 V, then press
up and release the TRIGGER COUPLING switch.
s. Disconnect the test setup.
CAL 04—CH 2 DELAY ENABLE/DISABLE
4. Check/Adjust CH 2 Delay Enable/Disable.
a. Scroll to CAL 04.
b. Press up and release the TRIGGER COUPLING
switch to initiate the routine.
c. CHECK— Readout alternately indicates “ENABLED”
and “DISABLED” each time the TRIGGER COUPLING
switch is pressed up and released.
d. Leave the readout display indicating “ENABLED”.
Press and release the A/B TRIG button to exit the routine.
e. Connect a 100 kHz, positive-going signal from the
Calibration Generator fast-rise output to the CH 1 OR X and
CH 2 input connectors via a 50 ft BNC cable, a 5X attenua
tor, and a Dual-Input Coupler.
f. Set:
VERTICAL MODE
CH 1 and CH 2
VOLTS/DIV
CH 1 and CH 2
Input Coupling
CH 1 and CH 2
A and B SEC/DIV
TRIGGER
SOURCE
MODE
COUPLING
SLOPE
g. Set the generator amplitude for a 3- to 5-division dis
play amplitude. Use the CH 1 and CH 2 POSITION controls
to vertically overlay the traces near the center of the grati
cule area.
h. Set the Horizontal POSITION control to set the rising
edge of the signal near the center vertical graticule line.
i. Press the X I0 MAG button in to obtain a magnified
display.
j. Pull out the B SEC/DIV knob.
k. CHECK— Readout indicates “CH 2 DLY— ADJUST
A” and that the A control will move the leading edge of the
CH 2 trace at least 1 division to either side of the CH 1
trace.
I. ADJUST—A control to superimpose the leading
edges.
m. Push in the B SEC/DIV knob.
N O TE
if the CH 2 Delay Adjust feature is to be disabled for
normal instrument use, perform the following steps;
otherwise, proceed to step r below.
n. Reenter the Diagnostic Monitor by pressing the AV
and At buttons simultaneously (hold them in), then press
and hold the TRIGGER SLOPE button. Release the buttons
after about 1 second.
o. Scroll to CAL 04.
p. Press up and release the TRIGGER COUPLING
switch until the readout indicates “DISABLED.”
q. Press and release the A/B TRIG button to return to
normal operating mode.
r. Return the CAL/NO CAL jumper to the NO CAL posi
tion and disconnect the test setup.
On (buttons in)
10 mV
50
9.
DC
5 ns (knobs locked)
CH 1
AUTO LVL
DC
+ (plus)
5-15
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...