2. Adjust TRACE ROTATION (R975) and Y-Axis
Alignment (R1848).
N OTE
If the previous step was not performed, first setup the
Initial Control Settings a t the beginning o f the C R T
adjustments, then proceed as follows.
a. Set:
A and B SEC/DIV
50 /is (knobs locked)
INTENSITY
As required for a well de
fined trace
At
(At readout)
INTENSITY
As required for well de
fined vertical cursors
b. Using the CH 2 Vertical POSITION control, align the
trace with the center horizontal graticule line.
c. Position one of the At cursors to the center vertical
graticule line using either the A or the AREF OR DLY POS
control.
d. ADJUST—TRACE ROTATION (R975 on the front
panel) to align the trace with the center horizontal graticule
line.
e. ADJUST—Y-Axis Alignment (R1848) to align the At
cursor with the center vertical graticule line.
f. Repeat parts d and e as necessary for the best aligned
display.
3. Adjust Geometry (R1870).
N OTE
If the previous step was not performed, first setup the
Initial Control Settings a t the beginning o f the C R T
adjustments, then proceed as follows.
a. Set:
Input Coupling
CH 2
50
Q
DC
Off (no readout)
Adjustment Procedure— 2445 Service
b. Connect 10
ms
time markers from the Time-Mark
Generator to the CH 2 input connector via a 50 Q BNC
cable.
c. Use the Horizontal POSITION control to align the time
markers with the vertical graticule lines. Use the CH 2 POSI
TION control to align the base of the signal with the bottom
graticule line.
d. Set the CH 2 VOLTS/DIV switch for at least a 6-divi
sion vertical display.
e. Use the CH 2 POSITION control to set the tops of the
time markers to graticule center.
f. Set the CH 2 VOLTS/DIV switch one position clock
wise to overscan the display.
g. ADJUST—Geometry (R1870) for minimum curvature
of the time markers across the entire graticule.
h. Disconnect the test setup.
4. Adjust Edge Focus (R1864).
N O TE
If the previous step was not performed, first setup the
Initial Control Settings at the beginning o f the C R T
adjustments, then proceed as follows.
a. Set:
Input Coupling
CH 2
50
DC
VOLTS/DIV
CH 2
0.2 V
INTENSITY
Midrange
b. Connect a 50 kHz, 8-division signal from the Primary
Leveled Sine-Wave Generator to the CH 2 input connector
via a 50 Q BNC cable.
c. Center the display on the graticule.
d. ADJUST— Edge Focus (R1864), FOCUS (front-panel
control), and ASTIG (R977, front-panel preset) for the most
uniform focus over the entire display.
e. Disconnect the test setup.
AV and At
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...