Operating Information—2445 Service
Figure 2-1. Line selector switch, line fuse, and detachable power cord.
INSTRUMENT COOLING
To prevent instrument damage from overheated com
ponents, adequate internal airflow must be maintained at
all times. Before turning on the power, firs t verify that both
the air-intake holes on the bottom of the cabinet and the
fan-exhaust holes in the rear panel are free of any obstruc
tion to airflow.
START-UP
The 2445 automatically performs power-up tests each
time the instrument is turned on. The purpose o f these
tests is to provide the user w ith the highest possible con
fidence level that the instrument is fu lly functional. If no
faults are encountered, the power-up tests normally w ill be
completed in under five seconds, after which the instru
ment w ill enter the normal operating mode. A failure of
any of the power-up tests w ill be indicated by either a
flashing TR IG 'D indicator on the instrument fro n t panel or
a bottom-line readout on the c rt in the form : TEST X X
F A IL Y Y (where XX is the test number and YY is the
failure code o f the failed test).
If a failure of any power-up test occurs, the instrument
may still be usable fo r some applications. To put the
instrument into the operating mode after a power-up test
failure, press the A/B TRIG button. If the instrument then
functions fo r your particular measurement requirement, it
may be used, but refer it to a qualified service technician
fo r repair o f the problem at the earliest convenience.
Additional information on the power-up tests may be
found in the "Maintenance" section o f this manual and in
Appendix A o f the Operators Manual. Consult your service
department, your local Tektronix Service Center, or nearest
Tektronix representative if additional assistance is needed.
REPACKAGING FOR SHIPMENT
If this instrument is to be shipped by commercial
transportation, it is recommended that it be packaged in
the original manner. The carton and packaging material in
which your instrument was shipped to you should be saved
and used fo r this purpose.
If the original packaging is u n fit fo r use or is not avail
able, repackage the instrument as follows:
1. Obtain a corrugated cardboard shipping carton
having inside dimensions at least six inches greater
than the instrument dimensions and having a carton
test strength of at least 275 pounds.
2-2
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...