Maintenance— 2445 Service
CORRECTIVE MAINTENANCE
INTRODUCTION
Corrective maintenance consists of component replace
ment and instrument repair. This part of the manual de
scribes special techniques and procedures required to
replace components in this instrument. If it is necessary to
ship your instrument to a Tektronix Service Center for repair
or service, refer to the “Instrument Repackaging Instruc
tions” at the end of this section.
MAINTENANCE PRECAUTIONS
To reduce the possibility of personal injury or instrument
damage, observe the following precautions.
1.
Disconnect the instrument from the ac power source
before removing or installing components.
2. Verify that the line-rectifier filter capacitors are dis
charged prior to performing any servicing.
3. Use care not to interconnect instrument grounds
which may be at different potentials (cross grounding).
4. When soldering on circuit boards or small insulated
wires, use only a 15-watt, pencil-type soldering iron.
OBTAINING REPLACEMENT PARTS
Most electrical and mechanical parts can be obtained
through your local Tektronix Field Office or representative.
However, many of the standard electronic components can
usually be obtained from a local commercial source. Before
purchasing or ordering a part from a source other than
Tektronix, Inc., please check the “Replaceable Electrical
Parts” list for the proper value, rating, tolerance, and
description.
N O T E
P hysical size an d shape o f a com ponent m ay affe c t
instrum ent perform ance, particularly a t high frequen
cies. A lw ays use direct-replacem ent com ponents, un
less it is know n th at a substitute w ill n o t degrade
instrum ent perform ance.
Special Parts
In addition to the standard electronic components, some
special parts are used in the 2445. These components are
manufactured or selected by Tektronix, Inc. to meet specific
performance requirements, or are manufactured for
Tektronix, Inc. in accordance with our specifications. The
various manufactures can be identified by referring to the
“Cross Index-Manufacturer’s Code number to Manufactur
er” at the beginning of the “Replaceable Electrical Parts”
list. Many of the mechanical parts used in this instrument
were manufactured by Tektronix, Inc. Order all special parts
directly from your local Tektronix Field Office or
representative.
Ordering Parts
When ordering replacement parts from Tektronix, Inc.,
be sure to include all of the following information:
1. Instrument type (include modification or option
numbers).
2. Instrument serial number.
3.
A description of the part (if electrical, include its full
circuit component number)
4. Tektronix part number.
MAINTENANCE AIDS
The maintenance aids listed in Table 6-11 include items
required for performing most of the maintenance procedures
in this instrument. Equivalent products may be substituted
for the examples given, provided their characteristics are
similar.
6 - 1 8
Summary of Contents for 2445
Page 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Page 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Page 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Page 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Page 210: ...3829 58 Figure 9 4 2445 block diagram ...
Page 214: ......
Page 217: ......
Page 219: ...2445 382 72 ...
Page 222: ...2445 ...
Page 231: ...A 1 t C t t F t G t H t ...
Page 233: ......
Page 236: ......
Page 238: ...2445 392 1 75 ...
Page 244: ......
Page 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Page 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Page 253: ......
Page 263: ... 0 2445 J8 i S ...
Page 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Page 275: ......
Page 278: ......
Page 281: ......
Page 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Page 286: ...2445 3823 82 ...
Page 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Page 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Page 300: ......
Page 304: ...2 R E TU R N T O 1 ...
Page 305: ...ERROR MESSAGE DIAGNOSTICS ...
Page 306: ...ERROR MESSAGE DIAGNOSTICS ...
Page 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Page 308: ...RETURN TO ...
Page 309: ...FRONT PANEL TROUBLESHOOTING ...
Page 310: ...FRONT PANEL TROUBLESHOOTING ...
Page 311: ...2445 Service 3829 90 ...
Page 316: ...R E TU R N T O v 1 y ...
Page 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Page 323: ......
Page 324: ...2445 Service 3829 85 ...
Page 325: ......
Page 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Page 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Page 330: ......
Page 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Page 338: ......
Page 346: ...12 2445 SERVICE ...
Page 347: ...2445 SERVICE ...
Page 348: ...2445 SERVICE ...