Super Systems Inc.
Page 76 of 107
Color Touch Screen Data Logger Operations Manual
SDS Data Tags
SDS Data tags are created so that all of the template information and survey data can
automatically be used for the report process. Data that is captured during the survey is
summarized to provide the overall results for the report. This can be in the form of a graph,
tabular data, text, etc. Data tags have been specifically created to address the uniformity
requirements for AMS and other standards. We have tags that address overshoot, minimum TC
with value, maximum TC with value, trend data, tabular data, etc. An example of an output from
the data tags can be seen by opening the Example report from the SDS Reporter provided with
the installation. The only data not generated from data tags is the tabular print out of the actual
temperatures. This data is generated from the
Survey TC Data
option and will print all data
points that are displayed on the graph between the “start” and “stop” selected by the user in the
survey.
In the figure below (Output From Data Tags), you can see an example of the data tags for the
graph, survey setpoint, minimum and maximum TCs with values and deviation. The data tags
used to generate the information below are:
<SDS:GRAPH600x400>
TC Survey Summary
Temperature Setpoint: <SDS:SP>°#
Minimum TC number:
#mtcn#
Maximum TC number:
#xtcn#
Minimum TC Value:
#mtcv#
Maximum TC Value:
#xtcv#
Min deviation from setpoint: #md#
Max deviation from setpoint: #xd#