Super Systems Inc.
Page 53 of 107
Color Touch Screen Data Logger Operations Manual
Figure 66 - Spool Correction screen
All values displayed to the users are corrected
. That means if there is a defined offset for the
TC Spool and a channel offset, the user would be shown the net result of those offsets into the
value displayed.
There are checkboxes for a possible forty thermocouples, but only the active thermocouples will
have checks in the checkboxes. Clicking on the “Define Control TC” link will allow the user to
select the TC or TCs that will be the control TC(s) (Figure 67). The operator can select any of the
thermocouples to be the control TC(s) and provide an optional description for that TC.
Figure 67 - Define Control TC screen
Clicking on the Accept button will set the selected control TC(s). Once a control TC(s) has been
selected, the TC’s number will be in
Red
on the “Active Thermocouples” section of the
Thermocouples tab.
The thermocouple(s) that is identified as the control TC will not be used for
the TC summary when identifying hottest and coldest channels.
Clicking on the “View Offsets used in Survey” link will display a list of the offsets that were used
on the survey (
Figure 68
).
All values that are displayed to the users are corrected values.
That
means the Offset defined for these TCs have already been incorporated into the value the
operator would see. IF there is a value for the TC Spool Correction Factor, that value will also
be incorporated providing a corrected view for the user.