Color Touch Screen Data Logger Operations Manual
The information for the comparison check is generated from several of the Report Properties
fields.
Report Field
Location in Tabs
Time
Chart Start Time
Comparison Test T/C
Spool Number
Middle Left of Comparison
Check
Comparison T/C
Number
Top Right of Comparison
Check
Comparison T/C Wire
Correction Factor
Top Right of Comparison
Check
Comparison
Instrument Correction
Factor
Thermocouples
Comparison T/C
Temperature Readings
Records the temperature
of the Comparison T/C
Adjacent T/C Number
for Comparison Check
Top Right of Comparison
Check
Adjacent T/C Wire
Correction
Thermocouples, click View
Offsets Used in Survey
button*
Adjacent Instrument
Correction Factor
Thermocouples, click View
Offsets Used in Survey
button
Adjacent T/C
Temperature Readings
for Comparison Check
Records the temperature
of Adjacent T/C
CC Difference
Difference Between
Comparison T/C and
Adjacent T/C temperatures
Accept/Reject Criteria =
Tolerance of Furnace
Set on Comparison Check
Page, Middle Right
Accept/Reject
Dependent on A/R Criteria
Table 2 - Location of Report Fields
*The offsets for the T/C Wire Correction can be adjusted solely for the purpose of the
comparison by checking the box on the Comparison Check page labeled Use Custom Wire
Corrections.
A header text and a body text can be put in the report. There are Preview Report and Print
Report buttons in the top left hand corner of the Comparison Check page.