764333-766
WARNING – ELECTRIC SHOCK HAZARD:
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Unplug surgical table from facility electrical supply and remove battery fuse before servicing. Always
follow appropriate electrical safety-related work practice standards.
Failure to keep all personnel and equipment clear of the table before actuating any inertia-driven or
power-driven movement could result in table damage.
Hang the hand control from side rail of the table when not in use to avoid possible damage to the control.
When cleaning/disinfecting table, thoroughly read the cleaning fluid directions for use and follow all
directions and cautions as shown.
Do not spray cleaning fluid into electric receptacles and avoid spraying directly on emergency backup
buttons or into clearance space. Spray or drippage may settle onto electric circuits inside table causing
corrosion and loss of function.
When moving the table to or from point of use, roll it carefully at moderate speed and only over smooth
floors. Maximum floor clearance is 1/4" (6.4 mm). Avoid door and elevator jambs and obstructions
greater than 1/4" (6.4 mm). If necessary, lift table over obstructions, onto trucks, etc. Lift table evenly
and only by the table base. If transporting articles (including accessories) on top of the table, ensure all
articles are properly secured to prevent falling off tabletop.
DO NOT use a forklift to move the uncrated
table
.
Cleaning procedures requiring articulation of the table should be performed only by persons familiar
with table operation.
Route the hand control cord clear of any pinch points where cord could be damaged.
Use of incorrect hydraulic oil may severely damage the table and/or cause table malfunction. Contact
STERIS for proper hydraulic oil.
After performing cleaning procedures, ensure pads and tabletop are completely dry before reinstalling.
Failure to do so can result in damage to tabletop and pads.
Appropriate components of this surgical table have been tested and found in compliance with IEC
60601-1-2:2001, Medical Electrical Equipment - Part 1; General Requirements for Safety; Electromagnetic
Compatibility (EMC). There is, however, a potential for electromagnetic or other interference between
this equipment and other devices. Should user experience interference, relocate this device or minimize
the use of the affected equipment while this device is in operation.
Do not pull the hand control cord when disconnecting hand control from table. This can result in damage
to the hand control.
Possible EMC increased EMISSIONS or decreased IMMUNITY. Do not use accessories or replacement
parts not listed in the Operator or Maintenance Manuals.
Static electricity can damage microprocessor controls. Indiscriminate probing of circuits or improper
connections may result in immediate or delayed electrical failure. If the electronics are to be accessed,
use a personal grounding device. Exercise care in all activities involving the electronic circuitry.
Use care when separating the bearing assembly from each side of the table as the sliding tabletop side
member is now free-floating when the bearing assembly is removed. Avoid bumping or misaligning
the slide potentiometer pulley.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......