6-2
150832-639
Operator Manual
Troubleshooting
* Service charges may be incurred. Consult your warranty for details.
** Use optional foot control or auxiliary (override) control system if hand control is faulty [refer to
].
Table 6-1. Troubleshooting Chart
SITUATION
ACTION
Hand Control does not activate when table is plugged
into electrical outlet and any pushbutton (except
STOP
) is pressed .
Main battery fuse blown – replace.
Faulty power cord – replace.
Hand control not plugged in properly – correct.
Faulty hand control** – replace.
Call STERIS.*
Table remains on battery power even when main
power cord is plugged into facility power.
Incorrect power cord connection – correct.
Faulty power cord – replace.
Main fuse blown – replace.
Call STERIS.*
No functions using hand control.
Hand Control not plugged in properly – correct.
No Power – table deactivates after 30 minutes of non-
use. Press any button on hand control (except
STOP
).
Faulty hand control** – replace.
Call STERIS.*
Table does not tilt right/left using hand control.
Hand control not on – Press any button on hand
control (except
STOP
).
Faulty Hand control** – replace.
Call STERIS.*
Table does not Trendeleburg/Reverse Trendeleburg
using hand control.
Hand control not on – Press any button on hand
control (except
STOP
).
Faulty Hand Control** – replace.
Call STERIS.*
Table does not raise/lower using hand control.
Hand control not on – Press any button on hand
control (except
STOP
).
Faulty hand control** – replace.
Call STERIS.*
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......