2-8
150832-639
Operator Manual
Important User Information
2.5.4 Weight
•
Table: 560 lb (250 kg)
Maximum Patient Weight (patient in normal orientation): 1100 lb
(499 kg) without patient posturing or tabletop slide (except raise/
lower); 1000 lb (454 kg) full patient posturing (except tabletop
slide); 600 lb (272 kg) full patient posturing (including tabletop
slide).
2.5.5 Power Requirements
Line Power Input:
•
100-240 Vac, One-Phase, 50/60 Hz, 4.0 Amp
NOTE: Each table is shipped from the factory configured to the
electrical requirement specified on the factory order. If this electrical
configuration needs to be changed in the field,consult STERIS for the
needed procedure and required materials.
2.5.6 Environmental
Conditions
Temperature:
32-122°F (0-50°C)
Relative Humidity:
20-80% RH
2.5.7 Image Amplification
Coverage
Figure 2-5. STERIS 4085 General Tabletop Dimensions (Typical)
81” (2055mm)
22”
(550mm)
13”
(325mm)
22”
(570mm)
22”
(550mm)
24”
(610mm)
3” (80mm)
13”
(325 mm)
22”
(570 mm)
22”
(550 mm)
24”
(610 mm)
3” (80 mm)
81” (2055 mm)
Dimensions are inches (mm). SAE measurements
are approximate based on metric dimensions.