4-6
764333-766
Batteries require recharging on a periodic basis
depending on frequency of table usage. Low or
discharged battery conditions are indicated by HALF
FILLED or EMPTY BATTERY ICON on the Hand
Control and some or all green diode bars on Power
Panel LED Display are OFF.
NOTE: Fully charged batteries enable between 100
and 150 procedures (approximately three weeks)
without recharging. Note the charge/discharge icon on
the LED Display (see Figure 4-1) of Power Panel.
Lead acid batteries last longer if not fully discharged.
Therefore, to obtain the longest life and capacity from
your STERIS 4085 General Surgical Table batteries,
always connect ac power cord to table base and plug
into an appropriate ac receptacle as often as possible,
and as long as possible, with the Main Power Switch
set to
ON
(lighted green). If this is not always possible,
recharge batteries at the following times:
NOTE: When two green diode bars on Power Panel
LED Display are OFF, batteries are half charged. This
level of power enables normal table operation without
damaging the batteries. However, when the last green
diode bar is OFF, batteries must be recharged without
further delay or battery damage occurs. If all diodes
are OFF, batteries are completely drained and the table
can only operate on main facility power.
• When table is first put into service.
• Every 15 days when table is in normal service; more
often if usage demands.
• Whenever HALF FILLED Battery ICON appears on
LCD Status Control Indicator.
• If table remains in extended storage for longer than
three months, batteries must be charged every three
months.
NOTE: If batteries do not charge, refer to S
, for possible causes and corrective
actions.
Recharge batteries as follows:
1. Connect ac power cord to table base and plug into
an appropriate ac receptacle.
2. Turn Main Power Switch to ON (lights green).
3. Allow
a minimum of 12 hours
for full battery
charge.
4. Verify green FILLED BATTERY ICON appears on
Hand Control and disconnect ac power.
NOTE: Depending on software version, if battery
voltage drops below either 19 V,
Plug Mains
alarm
appears or below 23 V,
Lo Batt
alarm appears. Refer
to S
, if either alarm
appears.
4.5 CHANGE FUSES
4.5.1 Battery
Install 16 A battery fuse (F8) into table base as follows
(see Figure 4-2):
1. Turn power
OFF
.
2. Unplug power cord from facility power.
3. Locate correct battery fuse.
4. Unscrew and remove black fuse cap.
5. Insert fuse in cap.
6. Insert fuse (with cap) into fuse holder assembly in
table base.
7. Push in fuse and cap and turn a quarter turn to lock
assembly into table base. Pull on cap to ensure fuse
is seated and locked properly.
8. Ensure table operates correctly on battery power.
9. Turn power
ON
.
NOTE: If table is going to be stored for more than
three months, remove fuse to limit battery discharge.
4.5.2 Main Control
Two replaceable fuses (F6 and F7, 4 Amp) are located
in a cartridge beside the ac input in the table base. If
one or both of the fuses are blown by a fault condition,
replace as follows (see Figure 4-2):
1. Turn power
OFF
.
2. Disconnect ac power cord from wall receptacle.
3. Pry cartridge out with a small screwdriver to access
fuses.
4. Remove blown fuse(s) and replace. See
for correct part number of fuses.
5. Push cartridge back into connected position.
6. Remove battery fuse (see
7. Reconnect ac power cord to wall receptacle.
8. Turn power
ON
.
9. Verify INTELLIPOWER dual power system
indicates table is operating on ac power (green LED
must be ON).
10. Ensure table functions properly.
11. Return battery fuse.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......