4-4
764333-766
4.3.2 After Each Usage
After each use of the STERIS 4085 General Surgical
Table, clean/disinfect as follows:
NOTE: Ensure all protective covers are installed over
any open receptacles.
1. Remove gross soil with a disposable cloth and place
used cloth in an appropriate biohazardous waste
disposal container.
2. Clean tabletop and pads as follows:
a.
Articulate tabletop to level and place at
comfortable working height.
b.
Gently "peel" pad away from tabletop fastening
strips and place on another table or flat surface.
c.
Hold can 6-8" (152-203 mm) from surface and
spray cleaning fluid liberally on top and sides
of pads. (It is recommended to clean only one
pad at a time.) Carefully avoid any pad seams.
NOTE: Do not use any oxidizing products with
chloric derivative base such as bleach, or
oxygenic derivatives such as peracetic acid.
d.
Wipe sprayed surfaces with a clean, lint-free
cloth dampened with water to remove cleaning
fluid (dampening cloth minimizes streaking).
e.
Wipe cleaned surfaces again with clean, damp,
lint-free cloth to remove residue.
f.
Wipe cleaned surfaces again with clean, dry,
lint-free cloth to remove all moisture.
g.
Repeat
Steps c, d, and e
for other pads. Repeat
Steps c, d, e, and f
for tabletop surfaces exposed
when pads are removed. Thoroughly dry
bottom of pads and tabletop surfaces.
h.
Place pads back onto tabletop by aligning with
sides of table and pressing into place on
fastening strips.
3. Raise table to maximum height to access lower
surfaces.
4. Clean yoke, shrouds and entire base surface as
follows:
a.
Hold can 6-8" (152-203 mm) from surface and
spray cleaning fluid liberally on yoke and
shrouds. Avoid spraying in direction of
connectors.
b.
Wipe sprayed surfaces with a clean cloth
dampened with water to remove cleaning fluid
(dampening cloth minimizes streaking).
c.
Wipe cleaned surfaces again with a clean,
damp, lint-free cloth to remove residue.
d.
Repeat
Steps a, b, and c
for base surface.
5. Unplug hand control. Note hand control plug is
locked, do not simply pull on plug.
6. Clean hand control as follows:
NOTE: Do NOT immerse hand control in liquid.
a.
Hold can 6-8" (152-203 mm) from surface and
spray cleaning fluid liberally on hand control.
b.
Wipe sprayed surfaces with a clean cloth
dampened with water to remove cleaning fluid
(dampening cloth minimizes streaking).
c.
Wipe cleaned surfaces again with a clean,
damp, lint-free cloth to remove residue.
d.
Wipe cleaned surfaces again with clean, dry,
lint-free cloth to remove all moisture.
7. If necessary, brush tabletop and pads fastening
strips with a non-metalic brush.
4.3.3 End-Of-Day Cleaning Procedure
At the end of each day, perform the cleaning
procedures as outlined in
4.3.4 Weekly Cleaning Procedure
After each weekly use of the Cmax Surgical Table,
clean/disinfect table as follows:
1. Perform
Steps 1 through 4
.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
• When cleaning/disinfecting table,
thoroughly read the cleaning fluid
directions for use and follow all
directions and cautions as shown.
• Cleaning procedures requiring
articulation of the table should be
performed only by persons familiar with
table operation.
• Do not spray cleaning fluid into electric
receptacles and avoid spraying directly
on emergency backup buttons or into
clearance space. Spray or drippage may
settle onto electric circuits inside table
causing corrosion and loss of function.
• After performing cleaning procedures,
ensure pads and tabletop are completely
dry before reinstalling. Failure to do so
can result in damage to tabletop and
pads.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......