1-2
150832-639
Operating Manual
Safety Precautions
WARNING – PERSONAL INJURY HAZARD:
WARNING – INSTABILITY HAZARD:
WARNING – PINCHING AND TIPPING HAZARD:
Failure to keep the patient properly secured with the patient safety straps at all times could result in
death or serious injury.
The floor locks must be engaged at all times when a patient is on the table. Failure to engage the floor
locks before placing a patient on the table or disengaging the floor locks while a patient is on the table
could result in the table rolling unexpectedly during the procedure.
Keep hands and feet clear of the unloading platform and the table base when unloading the table.
Serious personal injury could result.
Section 4.1,
Steps 1-6
in "Operating Table" must be executed before patient is transfered to the table.
Execution of
Steps 1-6
with the patient on the table could result in injury to both the patient and the
operating room staff.
Healthcare professionals must ensure patients are positioned and monitored so as to prevent
compromising respiration, nerve pathways or circulation.
When installing any table accessory, check for correct attachment and tighten securely (if appropriate).
Do not use a worn or damaged accessory. Before using any accessory, ensure it has been installed
properly.
Patient must be secured to the table in accordance with recommended positioning practices.
Unanticipated patient or table movement could result in death or serious injury.
When cleaning/disinfecting table, do not use phenolics which may cause patient skin burns if
inadequately rinsed off, or alcohol, which does not have sufficient cleaning/disinfecting properties.
The table leg section is designed to support the patient legs.
Do not put upper body weight on the leg
section
.
If the integrity of the external protective ground installation or arrangement is in doubt, operate the table
from its internal power source.
If an antistatic path is necessary, STERIS recommends antistatic pads (specifically developed for this
table) in direct contact with the patient. Table must also be positioned on antistatic floor or connected to
equalization device (equipotential connector).
Do not remove table leg section if patient is not properly positioned. Patient
'
s torso and buttocks must
be firmly held in place by table back and seat sections and legs supported by leg supports.
Possible patient or user injury, as well as table or accessory failure, may result from using STERIS table
accessories for other than their stated purpose – or from using accessories manufactured and sold by
other companies on STERIS surgical tables.
Patient injury may result if the operator of this table is not completely familiar with the controls for patient
positioning and table operation.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......