4-6
150832-639
Operator Manual
Operating Instructions
4.2.2 Hand Control LCD
Display Description
The Hand Control is equipped with a LCD (Liquid Crystal Display)
Display to assist in the table operation (see Figure 4-4). The LCD
Display indicator descriptions are as follows:
NOTE: Numbers in ( ) refer to items shown in Figure 4-4.
LCD Display (1):
displays any control system messages to the user
about table operation.
LCD Status ICONs (2):
indicates the table is receiving power from
the table batteries (Battery ICON) or facility power (Plug ICON) and
whether the table floor locks are engaged or not (Lock or Open Lock
ICONS). Refer to
. This area may also indicate an
operational situation, see Maintenance (11). Refer to
, for more information.
Tabletop Positioning (3):
indicates what tabletop function is in
operation.
NOTE: Auto Limit Sensors (ALS) Warning indicates the tabletop has
been ordered into a configuration or has moved into a position that
may cause tabletop damage or patient injury. This feedback is from
the Auto Limit Sensor™ (patent pending) control device system. To
prevent a tabletop sectional conflict, or collision of tabletop with the
table or floor, ALS System immediately stops the movement in
process and the ALS LED lights. The control system alerts the user to
an alternative movement choice in the LCD Display (1).
IMPORTANT:
The ALS System suggested alternative movement is
not the only movement available.
Figure 4-4. Primary Hand Control LCD Display
4.3 Reverse Patient
Orientation
Some surgeries (e.g., shoulder) require the patient to be in reverse
positioning and the table to be operated in a reverse configuration.
Press
Reverse Orientation
button on hand control (see Figure 4-3)
and hold for two seconds, Yellow LED should light beside graphic
representation of Reverse Orientation. Press
Reverse Orientation
button again and hold for two seconds to cancel Reverse Orientation
function and green LED lights beside graphic representation of
Normal Orientation.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: The Auto
Limit Sensor (ALS) System
suggested alternative tabletop
movement does not take into
account the type of surgery
that may be in process or the
environment (e.g., other
equipment close by) and
might not be compatible with
the surgical procedure. To
avoid an action that could
result in death or serious
injury to the patient and the
operating room staff, and/or
serious equipment damage,
completely evaluate the
feasibility of the suggestion.
KEY:
1. LCD Display
2. LCD Status Indicators (Refer To
)
3. Tabletop Positioning
1
2
3
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......