5-5
150832-639
Operator Manual
Routine Maintenance
3.
Wipe caster with a cloth, dampened with water, to remove
cleaning fluid and debris. Ensure items such as suture, oils and
floor wax are removed.
4.
Perform
Steps 2 and 3
for remaining three casters.
5.5 Battery Charging
Procedure
STERIS 4085 General Surgical Table is equipped with INTELLIPOWER
®
dual power system. The table is powered either by facility power or by
internal batteries. If table is to be operated on battery power, note the
following:
IMPORTANT:
Batteries are recharged automatically as long as the
Main Power Switch is set to
ON
(lighted green) and the ac power is
connected to the STERIS 4085 General Surgical Table. The PLUG
icon appears on Power Panel LED Display (see Figure 5-1)
indicating ac power and diode bar indicates battery condition.
Bar graph description for 120 V tables:
•
When table is plugged into facility ac and Main Power Switch is
ON, diode bar shows battery charge state. This battery charge
state is also indicated by icon on primary hand control (refer to
Figure 4-4).
•
When table is on battery power and primary hand control is
active (on), diode bar shows battery charge state. This battery
charge state is also indicated by icon on primary hand control
(refer to Figure 4-4).
Bar graph description for 220 V tables:
•
When table is plugged into facility ac and Main Power Switch is
ON, diode bar shows battery charge state. This battery charge
state is also indicated by icon on primary hand control (refer to
Figure 4-4).
•
When table is on battery power and primary hand control is
active (on), diode bar is off. Battery charge state is indicated by
icon on primary hand control (refer to Figure 4-4).
Batteries require recharging on a periodic basis depending on
frequency of table usage. Low or discharged battery conditions are
indicated by HALF FILLED or EMPTY BATTERY icon on the Hand
Control (refer to Figure 4-4 and
) and some or all green diode bars on Power
Panel LED Display are OFF.
NOTE: Fully charged batteries enable between 100 and 150
procedures (approximately three weeks) without recharging. Note the
charge/discharge ICON on the LED Display (see Figure 5-1) of
Power Panel.
Lead acid batteries last longer if not fully discharged. Therefore, to
obtain the longest life and capacity from your STERIS 4085 General
Surgical Table batteries, always connect ac power cord to table
base and plug into an appropriate ac receptacle as often as
possible, and as long as possible, with the Main Power Switch set to
ON
(lighted green). If this is not always possible, recharge batteries
at the following times:
NOTE: When two green diode bars on Power Panel LED Display are
OFF, batteries are half charged. This level of power enables normal
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Repairs
and adjustments to this
equipment must be made only
by STERIS or STERIS-trained
service personnel.
Maintenance performed by
unqualified personnel or
installation of unauthorized
parts could cause personal
injury, result in improper
equipment performance,
invalidate the warranty, or
result in costly damage.
Contact STERIS regarding
service options.
WARNING – TRIPPING
HAZARD: Route the power
cord to the receptacle to
minimize being tripped over
by personnel in the area.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......