3-1
Installation Instaructions
Operator Manual
150832-639
Check the overall condition of the STERIS 4085 General Surgical
Table and any applicable accessories upon receipt to ensure
nothing was damaged during shipment. If damage appears to
have occurred, contact the shipping company and STERIS.
STERIS recommends that only STERIS-trained or STERIS-supervised
personnel attempt to uncrate/crate, install and perform maintenance
on the STERIS 4085 General Surgical Table. Uncrating, crating,
installation or maintenance by others may result in personal injury or
damage to the product. Any alteration of this equipment not
authorized or performed by STERIS will void the warranty, could
adversely affect proper table functioning, could violate national, state
and local regulations.
NOTE: A patient grounding post/potential equalization terminal (male
connector, DIN 42801) is provided. The mating female connector for
patient grounding is not furnished by STERIS.
IMPORTANT:
Before connecting the table to your ac power system,
ensure table is marked for your power system (100, 120, 220 or 230/
240).
Install the table as follows:
1.
Position table at desired location.
NOTE: The power cord is approximately 20 ft (6 m) long. Ensure
table is positioned properly.
2.
The STERIS 4085 General Surgical Table is equipped with
INTELLIPOWER
®
dual power system modes. If table is to be
operated on battery power, proceed to
Step 8
.
3.
If not already installed, install battery fuse into table base as
follows (see Figure 3-1):
a.
Locate correct battery fuse.
b. Unscrew and remove black fuse cap.
c.
Insert fuse in cap.
d. Insert fuse (with cap) into fuse holder assembly in table base.
e.
Push in fuse and cap and turn a quarter turn to lock assembly
into table base.
NOTE: If table is going to be stored for more than six months,
remove fuse to limit battery discharge.
4.
Plug power cord into appropriate socket on table base (see
Figure 3-2).
5.
Plug power cord into appropriate facility outlet.
WARNING – PERSONAL
INJURY HAZARD:
• Keep hands and feet clear
of the unloading platform
and the table base when
unloading the table.
Serious personal injury
could result.
• If the integrity of the
external protective ground
installation or arrangement
is in doubt, operate the
table from its internal
power source.
WARNING – EXPLOSION
HAZARD: Table must not be
used in the presence of
flammable anesthetics.
WARNING – TRIPPING
HAZARD: Route the power
cord to the receptacle to
minimize being tripped over
by personnel in the area.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Table is
factory set to operate at a
certain voltage. Use of any
other power supply could
result in serious personal
injury or table damage.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: When
moving the table to or from
point of use, roll it carefully at
moderate speed and only over
smooth floors. Maximum floor
clearance is 1/4" (6.4 mm).
Avoid door and elevator jambs,
and obstructions greater than
1/4" (6.4 mm). If necessary, lift
table over obstructions, onto
trucks, etc. Lift table evenly and
only by the table base. DO NOT
transport articles (including
accessories) on top of the table,
and DO NOT use a forklift to
move the uncrated table.
WARNING – PERSONAL INJURY HAZARD: If an antistatic
path is necessary, STERIS recommends the use of
antistatic pads (specifically developed for this table) in
direct contact with the patient. Table must also be
positioned on antistatic floor or connected to equalization
device (equipotential connector).
INSTALLATION INSTRUCTIONS
3
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......