4-1
Operating Instructions
Operator Manual
150832-639
4.1 Operating Table
Operate the STERIS 4085 General Surgical Table as follows:
1.
Review Pinch Point Warnings as presented in
of this manual.
2.
STERIS 4085 General Surgical Table is equipped with INTELLIPOWER
®
dual power system. If table is to be operated on battery power, proceed
to
Step 4
.
NOTE: For more information on the battery system, refer to
S
3.
Verify power cord is plugged into both the facility outlet and table
receptacle and Main Power Switch is in ON position (lighted green).
4.
For optimum performance, before attempting to operate table,
allow table to reach room temperature
(see Figure 4-1)
.
5.
After referencing
position tabletop to LEVEL.
NOTE: If the Hand Control AND Foot Control are installed, Hand
Control has priority. Also, the STERIS 4085 General Surgical
Table Head Section is manually adjusted as follows (if Head
Section is not installed, refer to S
Step 11
):
1) The head section may be positioned UP and DOWN 90°
from horizontal (level).
2) While holding head section securely in one hand, pull
release handle (spring loaded) located under right side of
head section (see Figure 4-2) and tilt head section to
desired angle (level in this case).
3) Release handle to lock head section at desired angle. Move
headrest slightly until ratchet mechanism locks headrest in
position.
WARNING – PERSONAL
INJURY HAZARD:
• Section 4.1, Steps 1 - 6 in
"Operating the Table" must
be executed before patient
is transfered to the table.
Execution of Steps 1 - 6
with the patient on the table
could result in injury to
both the patient and the
operating room staff.
• Healthcare professionals
must ensure that patients
are positioned and
monitored so as to prevent
compromising respiration,
nerve pathways or
circulation.
• Patient must be secured to
the table in accordance with
recommended positioning
practices. Unanticipated
patient or table movement
could result in death or
serious injury.
• If the integrity of the
external protective earth
conductor installation or
arrangement is in doubt,
operate the table from its
internal power source.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD:
• Table is factory set to
operate at a certain voltage.
Use
of any other power
supply could result in
serious personal injury or
table damage.
• Use of this table in the
presence of flammable
anesthetics could result in
serious personal injury or
table damage.
Figure 4-1. STERIS 4085 General Surgical Table (Typical)
OPERATING INSTRUCTIONS
4
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......