4-10
150832-639
Operator Manual
Operating Instructions
4.5.3 Return Backup Control
System
When situation has been corrected and normal table operation is
available, replace Backup Control System as follows (see Figure 4-5):
1.
Turn Backup Hand Control face down.
2.
Wind cable and place under controller.
3.
Slide controller and cord back into table base.
4.
Fold pedal back to original position.
4.6 Leg Section
Removal
The STERIS 4085 General Surgical Table leg section is designed to
be easily removed (Hi-Lock
™
locking mechanism) to enable
procedures requiring optional leg supports. Remove the leg section
as follows:
1.
Using hand control (see Figure 4-3), level leg section (bring to
horizontal).
2.
While holding leg frame, press both handles and simultaneously
pull backward (see Figure 4-7).
3.
Safely store leg section to prevent damage.
NOTE: When leg section is removed from the STERIS 4085 General
Surgical Table, the LCD Display shows LEG OUT if the LEG UP or
LEG DOWN hand control pushbutton is pressed.
Figure 4-6. Backup (Override) Hand Control
TREND
REVERSE
TREND
HEIGHT UP
HEIGHT DN
BACK UP
BACK DN
LEG UP
LEG DN
FLOOR
LOCK
STOP
CAUTION
Back-up mode
Auto Limit Sensors NOT available
USE WITH FOOT PUMP
SLIDE HEAD
SLIDE FOOT
KIDNEY UP
KIDNEY DN
FLOOR
UNLOCK
TILT LEFT
TILT RIGHT
Green LED
Green LED
WARNING – PERSONAL
INJURY HAZARD: Do not
remove table leg section if
patient is not properly
positioned. Patient's torso and
buttocks must be firmly held
in place by table back and seat
sections and legs supported
by leg supports.
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......