HRX-OM-S004
Chapter 5 Display and setting of various functions
5.20
Alarm customizing function
HRS Series
5-50
5.20.2 Setting and checking of the alarm customizing function
The table below shows the set items of the alarm customizing function and
the default settings.
Table 5.20-5 Default settings of the alarm customizing function (1/3)
Object alarm
Display
Content
Code
Alarm name
Change content
Default
setting
∗
1
Operation setting when
"Low level in tank" alarm
is generated
AL01
Low level in tank
Product operation
while the alarm is
being generated
A.RUN
Operation setting when
"Detection temp. for the
circulating fluid discharge
temp. rise" alarm is
generated
Product operation
while the alarm is
being generated
A.RUN
Threshold temperature
setting for "Detection
temp. for the circulating
fluid discharge temp. rise"
alarm
AL03
Circulating fluid
discharge temp. rise
Alarm threshold
40.0
o
C
(104.0
o
F)
(----)
Operation setting when
"Detection temp. for the
circulating fluid discharge
temp. drop" alarm is
generated
Product operation
while the alarm is
being generated
A.RUN
Threshold temperature
setting for "Detection
temp. for the circulating
fluid discharge temp.
drop" alarm
AL04
Circulating fluid
discharge temp. drop
Alarm threshold
1.0
o
C
(33.8
o
F)
(----)
Operation setting when
"Circulating fluid
discharge pressure rise"
alarm is generated
Product operation
while the alarm is
being generated
ASTP
Threshold pressure
setting for "Circulating
fluid discharge pressure
rise" alarm
AL08
Circulating fluid
discharge pressure
rise
Alarm threshold
0.55MPa
(80PSI)
Operation setting when
"Circulating fluid
discharge pressure
decrease" alarm is
generated
Product operation
while the alarm is
being generated
A.STP
Threshold pressure
setting for "Circulating
fluid discharge pressure
decrease" alarm
AL09
Circulating fluid
discharge pressure
drop
Alarm threshold
0.05MPa
(7PSI)
(----)
Operation setting when
"Communication error"
alarm is generated
Product operation
while the alarm is
being generated
OFF
Threshold monitoring time
setting for
"Communication error"
AL19 Communication
error
Alarm threshold
----
(30)
Setting function for
"Contact input signal 1"
detection
AL31
Contact input 1 signal
detection
Product operation
while the alarm is
being generated
A.STP
Setting function for
"Contact input signal 2"
detection
AL32
Contact input 2 signal
detection
Product operation
while the alarm is
being generated
A.STP
*1: Regarding the detail of the default setting, refer to “Table 5.20-1 Alarm setting and customizing
(1/4)~(4/4)”.
Summary of Contents for HRS090-A
Page 8: ...HRX OM S004 Contents HRS Series...
Page 20: ...HRX OM S004 Chapter 2 Name and Function of Parts 2 4 Operation display panel HRS Series 2 6...
Page 56: ...HRX OM S004 Chapter 3 Transport and Setting Up 3 5 Circulating Fluid Supply HRS Series 3 36...
Page 164: ......
Page 174: ......
Page 192: ...HRX OM S004 Chapter 8 Documents 8 9 Daily Check Sheet HRS Series 8 18...
Page 195: ......