HRX-OM-S004
Chapter 3 Transport and Setting Up
HRS Series
3.5
Circulating Fluid Supply
3-33
3.5.2
Fill of fluid without using auto fluid-fill function
To supply the circulating fluid without using automatic fluid-fill function, remove the upper
panel on the right side, and supply the fluid to the fluid-fill port on top of the tank.
1.
Remove the screws (7 pcs.) to remove upper panel on the right side.
Fig. 3-22 Removal of the right side panel
2.
Hold the handles and lift up the right side panel upper, and remove it.
Remove the tank lid on top of the tank.
Fig. 3-23 Removal of the right side panel and the lid of the fluid-fill port.