Data reset function
5-35
5.15 Data reset function
5.15.1 Data reset function
Resets the values set by user to the default values. Note that the
accumulated operating time will not be reset.
5.15.2 How to operate reset function
The table below shows the set item for data reset and the default setting.
Table 5.15-1 Set item for data reset
Display
Item
Contents
Default
Data reset
Resets all the data.
(Accumulated operation time will not be
reset.)
OFF
1.
Press and hold the [MENU] button for approximately 2 seconds.
Repeat pressing the button until the button-lock setting screen [
] appears on the
digital display.
PV
SV
PV
SV
PV
SV
Data reset
2.
Press the [SEL] button 13 times.
Data reset screen appears on the digital display.
PV
SV
3.
Select
with [
▲
] button or [
▼
] button, and press
[SEL] button to enter. Select
and press [SEL] button to enter. This resets all the data to the default, and the
display returns to the main display.
Table 5.15-2 Data reset
Set value
Explanation
Default
No data reset.
✓
Resets all the data.
Press and
hold [MENU]
key.
Press and
hold [MENU]
key.
This function resets all the set values.
Use caution when operating this function.
It is recommended to record the set data before resetting.
Summary of Contents for HRS090-A
Page 8: ...HRX OM S004 Contents HRS Series...
Page 20: ...HRX OM S004 Chapter 2 Name and Function of Parts 2 4 Operation display panel HRS Series 2 6...
Page 56: ...HRX OM S004 Chapter 3 Transport and Setting Up 3 5 Circulating Fluid Supply HRS Series 3 36...
Page 164: ......
Page 174: ......
Page 192: ...HRX OM S004 Chapter 8 Documents 8 9 Daily Check Sheet HRS Series 8 18...
Page 195: ......