HRX-OM-S004
Chapter 5 Display and setting of various functions
5.21
Communication function
HRS Series
5-70
Setting/Checking: Communication mode
1.
Press and hold the [MENU] key for approximately 2 seconds.
Repeat pressing the button until the communication mode setting screen [
]
appears on the digital display.
PV
SV
PV
SV
PV
SV
PV
SV
PV
SV
2.
Select a communication mode with [
▲
] key or [
▼
] key, and press [SEL] key to enter.
Table 5.21-2 Setting of communication mode
Set value
Explanation
Initial value
(Default setting)
Sets LOCAL mode.
(Operation and setting can be performed on the
operation panel.)
✓
Sets DIO mode.
∗
1
(The product is operated with contact
input/output.)
Sets SERIAL mode.
∗
2
(Operation and setting are performed by serial
communication.)
∗
1
:
When the contact input signal 1 is “external switch signal”, it is not possible to set to “DIO
mode”.
∗
2
:
When the serial protocol is “simple communication protocol 2” and the contact input 1 is
“external switch signal” or contact input 2 is “remote signal”, it is not possible to set to
“SERIAL mode”.
Setting/Checking: Serial communication protocol
3.
Press the [SEL] key once.
Setting screen of serial communication protocol appears on the digital display.
PV
SV
Press and
hold [MENU]
key.
Press and
hold [MENU]
key
Press and
hold [MENU]
key
Press and
hold [MENU]
key
Press and
hold [MENU]
key
Summary of Contents for HRS090-A
Page 8: ...HRX OM S004 Contents HRS Series...
Page 20: ...HRX OM S004 Chapter 2 Name and Function of Parts 2 4 Operation display panel HRS Series 2 6...
Page 56: ...HRX OM S004 Chapter 3 Transport and Setting Up 3 5 Circulating Fluid Supply HRS Series 3 36...
Page 164: ......
Page 174: ......
Page 192: ...HRX OM S004 Chapter 8 Documents 8 9 Daily Check Sheet HRS Series 8 18...
Page 195: ......