61 25 665 D3437
368
D3437
.
076.01.15
.
02 06.2012
9 Repair
Sirona Dental Systems GmbH
9.13 Cable
Service Manual GALILEOS
21. CAUTION! Push the green cable. The white cable is carried along. In
this way, you can prevent the sensitive fiber optic cable from being
damaged. Push both cables (together) into the cable track up to the
cable markings.
New cables do not have cable markings. Orientate yourself according
to the marking on the second (old) cable and make sure that both
cables protrude equally far out of the cable track once they have been
drawn in. Then make a mark on the new cable.
22. NOTICE! The cable ties should only fix the position of the cables.
They must not be tightened too much, otherwise fiber-optic cable L7
could be damaged.
Before installing the cable track in the stand, fix the cables at both
ends of the cable track with a cable tie.
23. Reinstall the cable track in the stand.
Installation of the cable track is performed in reverse order of the
removal.
9.13.2.4
Replacing cable L1 or grounding strap in cable track 1
Replacing cable L1 or grounding strap in cable track 1
The procedure for replacing cable L1 and the grounding strap is basically
analogous to the procedure described in chapter Replacing cable L7/