10 Maintenance
Sirona Dental Systems GmbH
10.6 Checking the grounding straps
Service Manual GALILEOS
Grounding strap on the image detector: GALILEOS
Grounding strap on the image detector
Check whether the grounding straps
have complete and firm contact
1. Perform a visual and “hands-on” inspection of the grounding straps to
ensure that they have complete and firm contact at the positions
marked.
If the grounding straps do not have correct contact:
ª
Fasten the grounding straps correctly.
2. Perform a visual inspection of the grounding straps for damage.
If the grounding straps are damaged:
ª
Replace the grounding straps.