02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E3 71 30/33/34/35/36/37/38/39/40/41
Error code
Description
Actions required
see
E3 71 30
E3 71 33
E3 71 34
E3 71 35
E3 71 36
E3 71 37
E3 71 38
E3 71 39
E3 71 40
E3 71 41
Up/down keys pressed on power-
on.
Light localizer key pressed during
power-on.
T key pressed during power-on.
R key pressed during power-on.
Service key actuated during power-
on.
Memory key actuated during power-
on.
Program selection key actuated
during power-on.
Radiation time key actuated during
power-on.
kV/mA key actuated during power-
on.
Patient symbol pressed during
power-on.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit ON, making sure that the
Multipad is not actuated during boot-up.
4. Repeat procedure and check function.
If the error occurs repeatedly ...
● Replace board DX71 or Multipad.