Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.6
X-ray tube unit
båÖäáëÜ
1. Loosen the two lower screws (A) (approx. 2 to 3 turns).
2. Push the diaphragm (B) upward and then toward the front.
3. Unscrew the two sleeves (C) with washers.
4. Remove the aluminum filter (D).
5. Remove the primary diaphragm (F).
6. Keep diaphragm (B) safe and store primary diaphragm (F) and
aluminum filter (D) in a safe place.
IMPORTANT: These components will be reused when attaching the
diaphragm unit to the new tube assembly.
"Type 2" diaphragm (serial no. 2201 and higher)
1. Loosen the two lower screws (A) (approx. 2 to 3 turns).
2. Push the diaphragm (B) upward and then toward the front.
3. Unscrew the two sleeves (C) with washers.
4. Remove the aluminum filter (D).
5. Keep diaphragm (B) safe and store aluminum filter (D) in a safe
place.
IMPORTANT: These components will be reused when attaching the
diaphragm unit to the new tube assembly.
NOTICE
Oil leakage!
The screws (E) from the ring (X-ray window) must not be loosened, as
otherwise oil may leak out.