General operating procedures
Service Manual GALILEOS
4.5
Using demo mode – operation without radiation release
båÖäáëÜ
4.5
Using demo mode – operation without radiation
release
Using demo mode – operation without radiation release
GALILEOS demo mode
For demo use, the "X-ray detector dummy for GALILEOS" (Order No. 61
19 007) should be used instead of the actual X-ray detector. For further
information, please refer to the instructions included with the dummy.
If the volume tomograph is to be presented as a demo unit at trade fairs
or exhibitions, it must be ensured that radiation release is blocked.
4.5.1
Switching on demo mode
Switching on demo mode
When operated in demo mode, the unit must not release any radiation.
For this reason, you must take the following safety measures:
1. Switch off the unit.
2. Remove the cover of the tube assembly.
3. Loosen the screws (A) and remove the cover plate (B).
DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.
A
B