9 Repair
Sirona Dental Systems GmbH
9.10 Head fixation device
Service Manual GALILEOS
2. Align the acquisition unit.
To do this, move the rotary knobs on the head fixation device to a
vertical position. Then align the acquisition unit so that the laser light
is shown in the middle of the vertical knobs of the head fixation device
and the bite holder.
3. Press the locking button (A) and push the flange (B) including the
head fixation device (F) back to the front so that the bore hole (C) is
exposed. Tighten the screw (D) firmly.
IMPORTANT: The acquisition unit should not be turned when
pushing the head fixation device back and tightening the screw.
4. Switch the unit off again.