4 General operating procedures
Sirona Dental Systems GmbH
4.4 Reading unit data
Service Manual GALILEOS
4.4.2
Reading Facescan unit data over the network
Reading FACESCAN unit data over the network
Opening the web dialog
1. In SIDEXIS Manager, start the
"Facescan Configuration"
application.
2. Click on the
"Facescanner in browser"
button.
ª
A password dialog box opens.
3. In the field
"User"
enter
"service"
.
4. In the field
"Password"
enter
"sirona"
.
ª
The
"Facescan Device Service "
web dialog opens.
Selecting the service dialog
Selecting the service dialog
➢ In the menu bar, select the menu item
"SERVICE"
.
ª
The
"Facescan Service Functions"
window opens.
Reading unit settings
1. Press the
"Get Device State"
button.
ª
A dialog box to read the unit settings opens.
2. Press the
"Press Here to Download"
button.
3. Save the archived unit settings to the hard disk.