61 25 665 D3437
122
D3437
.
076.01.15
.
02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E6 89 21
E5 89 22
E5 89 23
* As of X-ray detector serial number 5000, cable L28 can no longer be
replaced individually.
E7 89 25
E7 89 26
E7 89 27
Error code
Description
Actions required
see
E6 89 21
File system error.
● Check board DX89, replace if necessary. S. [ → 134],
Error code
Description
Actions required
see
E5 89 22
The power supply of the X-ray
detector does not respond or is the
wrong version.
● Check board DX89, replace if necessary. S. [ → 134],
● Check x-ray detector, replace if
necessary.
Error code
Description
Actions required
see
E5 89 23
Camera head in the X-ray detector
does not respond or wrong version.
● Check cable L27/L28*, replace if
necessary.
● Check board DX89, replace if necessary. S. [ → 134],
● Check x-ray detector, replace if
necessary.
Error code
Description
Actions required
see
E7 89 25
Image memory error.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and check function.
● Run software update.
● Check that the memory modules on board
DX89 are firmly fixed, replace board DX89
if necessary.
Error code
Description
Actions required
see
E7 89 26
Total exposure time was exceeded. ● Check cable L13 (CAN bus), replace if
necessary.
Error code
Description
Actions required
see
E7 89 27
At least 10 image segments are
defective.
● Check cable L13 (CAN bus), replace if
necessary.
If the error occurs repeatedly ...
● Check board DX89, replace if necessary.
● Replace the X-ray detector.