9 Repair
Sirona Dental Systems GmbH
9.6 X-ray tube unit
Service Manual GALILEOS
"Type 3" diaphragm
1. Loosen the two lower screws (A) (approx. 2 to 3 turns).
2. Push the diaphragm (B) upward and then toward the front.
Removing the defective X-ray tube
assembly
1. Loosen the four screws (A) and remove cover plate (B) incl. the cable
shielding (L3). CAUTION! Also pull cable L3 off connector X3 and the
ground cable off connector X304 on board DX6.
Tip: The ferrite core and cable shielding can remain on the cover
plate.
2. Detach cables L5, L6 and L15 from the rubber grommets and pull the
cables off of sockets J6 (L5), J2-J3 (L6) and J5 (L15) on board DX6.
Summary of Contents for galileos