61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
105
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E6 14 10/12 GALILEOS
Error code
Description
Actions required
see
E6 14 10
E6 14 12
Clock signals for sensor image
transfer not received on board DX1/
DX11 (...10).
Faulty detection of sensor image
transfer data signals on board DX1/
DX11; recurring (...12).
● Check cable L13 for crushed spots and
kinks and check connectors, repair or
replace if necessary.
● Check board DX89, replace if necessary. S. [ → 134],