61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
323
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.8
X-ray detector
båÖäáëÜ
6. NOTICE! Ensure that cable L13 is correctly laid in the nut (O) of the
X-ray detector.
Route the cable L13 as illustrated in the diagram, and attach the
cover shielding on the X-ray detector housing using the 2 clamps (D)
and screws (B).
Depending on the unit hardware version, the brackets on your unit
may differ slightly from those shown.
7. Secure the cable with clamp (E) and screw (C).
8. Re-attach the cover plate (A).
Summary of Contents for galileos
Page 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Page 2: ......
Page 387: ......